The Perkin-Elmer Spectrum One FTIR Spectrometer is capable of data collection over a wavenumber range of 370-7800 cm-1. It can be configured to run in single-beam, ratio, or interferogram modes. The best resolution is 0.5 cm-1. A mid-range Deuterated triglycine sulfate infrared detector processes signals with a 68340 integrated chip. The sample compartment is purgeable.
Instrument are interfaced to Perkin Elmer Spectrum software, which gives the user the ability to extract both qualitative and quantitative data from the spectrum, and generate custom reports.
The Perkin Elmer Spectrum One FT-IR spectrometer is a bench top instrument that provides all the following in one self contained unit:
- A large, purgeable, sample compartment. The Spectrum One can operate in ratio, single-beam, or interferogram mode.
- An optical system that gives you data collection over a total range of 7800 to 370 cm-1
(220 cm-1 with CsI beam splitter) with a best resolution of 0.5 cm-1
- A mid infrared detector – either DTGS (deuterated triglycine sulphate) or LiTaO3 (lithium tantalite) – as standard, and the option of using MCT (Mercury Cadmium Telluride) or PAS ( a photoacoustic detector)
- An electronic system based on the Motorola DSP56303 Digital Signal Processor and the Motorola 68340 Integrated Processor.
|